Page 1 - Reference System

User Manual Milli-Q Reference System

Page 2 - About Millipore

- 10 - Cabinet, Continued Details of the Main Display Item Description 1 LEDs 2 Main LCD 3 Main Keypad Right The use of the Right Keypad button

Page 3 - Legal Information

- 11 - Cabinet, Continued Up The use of the Up Keypad button is shown below. It is used to scroll up in a menu. Diagram 1 Action Diagram 2 Pres

Page 4 - Legal Information Continued

- 12 - Cabinet, Continued READY Mode – water quality values The READY Mode screen display is explained below. Diagram Explanation In this example

Page 5 - Safety Information

- 13 - Cabinet, Continued Port and cables The port and cable connections are explained below. Item Description Item Description 1 Feedwater por

Page 6 - Table of Contents

- 14 - Consumables Flow diagram The water flow through a System is shown here in a flow diagram. The various consumables are described below. Ite

Page 7

- 15 - Consumables, Continued Quantum Cartridge The Quantum Cartridge removes trace levels of ions and organic molecules. Item Description Quantum

Page 8 - Product Information

- 16 - Specifications and requirements Milli-Q® Water quality The water delivered from a POD Unit has the following characteristics. Parameter Spe

Page 9 - Cabinet

- 17 - Specifications and requirements, Continued Dimensions Materials of construction Please contact Millipore for a list of the Materia

Page 10 - Cabinet, Continued

- 18 - Specifications and requirements, Continued Feedwater The Feedwater requirements are listed here. Parameter Value Type Pre-treated water inc

Page 11

- 19 - Installation Overview Purpose This chapter explains how to install the System. Contents This chapter contains the following topics: Topic

Page 12

- 2 - About this User Manual Purpose This User Manual is intended for use with a Milli-Q® Reference Water Purification System. This User Manual is

Page 13

- 20 - Alarms generated during installation Overview During the installation of a System, certain Alarm messages are generated. This occurs because

Page 14 - Consumables

- 21 - POD Unit, tubing and power cord Separate POD Arm and Point Of Delivery Separate the POD Arm and the Point Of Delivery by cutting and removin

Page 15 - Consumables, Continued

- 22 - POD Unit, tubing and power cord, Continued Loop  Install the Inlet Strainer as shown here.  Connect one end of the feedwater tubing to the

Page 16 - Continued on next page

- 23 - POD Unit, tubing and power cord, Continued Alarm messages Because the System is starting without a Q-Gard Pack or a Quantum Cartridge instal

Page 17 - Materials of

- 24 - Installing the Q-Gard Pack Procedure Follow the steps below to install a new Q-Gard Pack. Step Action Diagram 1 Start in STANDBY Mode. NO

Page 18 - Feedwater

- 25 - Installing the Q-Gard Pack, Continued Procedure (continued) Step Action Diagram 5 Push the bottom of the Q-Gard Pack inwards. 6 Push th

Page 19 - Overview

- 26 - Installing the Quantum Cartridge Procedure Follow the steps below to install a new Quantum Cartridge. Step Action Diagram 1 Open the righ

Page 20

- 27 - Rinsing the System Procedure Follow the steps below to rinse the System. Step Action Diagram 1 Locate the clear tubing and the barbed fit

Page 21

- 28 - Rinsing the System, Continued Procedure (continued) Step Action Diagram 5 Push the POD Plunger all the way down and then release it to st

Page 22 - System Cabinet

- 29 - Installing a POD Pak Overview The installation of a POD Pak involves 2 steps. These are:  placing and flushing the POD Pak onto the POD Un

Page 23 - Check the date

- 3 - Legal Information Notice The information in this document is subject to change without notice and should not be construed as a commitment by

Page 24 - Installing the Q-Gard Pack

- 30 - Installing a POD Pak, Continued Registering (continued) Step Action Diagram 5 Press . 6 Press . 7 In this example, you choose Millipa

Page 25 - (continued)

- 31 - Registering UV Lamp timer Introduction The timer used for the UV 185 nm Lamp must be reset when the System is installed. If this is not done

Page 26

- 32 - Registering UV Lamp timer, Continued Procedure (continued) Step Action Diagram 5 Press . 6 Press . 7 Press . 8 Press 3 times on .

Page 27 - Rinsing the System

- 33 - Calibrating the Flowrate Introduction The Milli-Q Water flowrate should be calibrated when the System is installed. A 1 Litre graduated cyli

Page 28

- 34 - Calibrating the Flowrate, Continued Procedure (continued) Step Action Diagram 6 Place a 1 L Graduated Cylinder under the POD Unit. Press

Page 29 - Installing a POD Pak

- 35 - Software Overview Introduction The purpose of this chapter is to explain the various software used in the System. Contents This chapter con

Page 30 - Registering

- 36 - Software Map Map

Page 31 - Registering UV Lamp timer

- 37 - Standby Mode General information Purpose STANDBY mode is used primarily for:  maintenance actions, and  going to the Manager Menu. Displa

Page 32

- 38 - Description of Standby Menu Maintenance The Maintenance Menu is described below. Diagram 1 Diagram 2 Item Description Clean Strainer

Page 33 - Calibrating the Flowrate

- 39 - Description of Standby Menu, Continued Language Diagram 1 Diagram 2 Item Description Language Change the displayed language.

Page 34

- 4 - Legal Information Continued Trademarks Millipore, Elix, Q-Gard, Quantum, Milli-Q, BioPak, EDS-Pak, Millipak and Millipore Express are registe

Page 35 - Software

- 40 - Manager Menu Description How to enter See the Software Map at the beginning of this chapter. The map shows how to enter the Manager Menu. T

Page 36 - Software Map

- 41 - Description, Continued Set Points Diagram 1 Diagram 2 Item Description Strainer Frequency Change set points for controll

Page 37 - General information

- 42 - Description, Continued Setup Diagram 1 Diagram 2 Item Description Install Date Change the installation date. Buzzer Ch

Page 38 - Description of Standby Menu

- 43 - Ready Mode General information Purpose In READY Mode, water can be dispensed from the POD Unit. The System should be left in READY Mode mos

Page 39 - Language

- 44 - Description of Ready Menu Water Quality Diagram 1 Diagram 2 Item Description MQ Feed Quality View the feedwater quality (accessory) T

Page 40 - Description

- 45 - Description of Ready Menu, Continued Consumables Status Diagram 1 Diagram 2 Consumable Description Q-Gard UV 185 nm Lamp Q

Page 41 - Description, Continued

- 46 - Description of Ready Menu, Continued Service Tracking Diagram 1 Diagram 2 Item Description Installation Repair Service Contract Contra

Page 42

- 47 - Using the Milli-Q System Overview Introduction The purpose of this chapter is to explain:  various ways that water can be dispensed from th

Page 43 - Ready Mode

- 48 - Dispensing water Optimise Water Quality Product Water can be recirculated within the System before dispensing it. This helps optimised wate

Page 44 - Description of Ready Menu

- 49 - Dispensing water, Continued Using the POD Plunger To dispense water, press down on the POD Unit plunger while in READY Mode. Position Wa

Page 45 - Call Millipore

- 5 - Safety Information Statement Your Milli-Q System should be installed and operated according to the instructions in this manual. In particular

Page 46 - Information

- 50 - Dispensing water, Continued Volumetric dispensing (continued) Step Action Diagram 3 Select the desired volume of water to be delivered. P

Page 47 - Using the Milli-Q System

- 51 - Viewing water quality Procedure Follow the steps below to view the water quality. Step Action Diagram 1 Make sure the System is in READY

Page 48 - Dispensing water

- 52 - Viewing Operation Introduction VIEW OPERATION allows you to see the status of major components. Under the View Operation LCD, the following

Page 49 - Dispensing water, Continued

- 53 - Viewing Operation, Continued System Alerts An example Alert is shown here. This is an Alert that is currently being displayed on the botto

Page 50

- 54 - Viewing Consumable Status Introduction Consumables Status allows you to see information related to the various consumables. Procedure Follo

Page 51 - Viewing water quality

- 55 - Calling Millipore Introduction Call Millipore allows you to see contact information. A Millipore Service Representative can enter this infor

Page 52 - Viewing Operation

- 56 - Viewing Information Introduction INFORMATION allows you to view:  flow schematic information,  version information, and  serial number an

Page 53 - Viewing Operation, Continued

- 57 - Viewing Information, Continued Version The various versions for the System are shown here. This LCD shows the version used for various compo

Page 54 - Viewing Consumable Status

- 58 - Maintenance Overview Introduction The purpose of this chapter is to explain the common maintenance needed for a System. Contents This chapt

Page 55 - Calling Millipore

- 59 - Maintenance Schedule Consumables Item Maintenance needed When Q-Gard Pack Replacement Prompted to by an LCD message. Quantum Cartridge

Page 56 - Viewing Information

- 6 - Table of Contents Product Information ...

Page 57 - Version

- 60 - Replacing the Q-Gard Pack When The Q-Gard Pack should be replaced when one of the following Alarm or Alert messages is displayed.  Alarm me

Page 58 - Maintenance

- 61 - Replacing the Q-Gard Pack, Continued Removing (continued) Step Action Diagram 4 Remove the used Q-Gard Pack. 5 The System will indicate

Page 59 - Maintenance Schedule

- 62 - Replacing the Q-Gard Pack, Continued Placing (continued) Step Action Diagram 3 Push the Pack Locking Handle down. Close the left door.

Page 60 - Replacing the Q-Gard Pack

- 63 - Replacing the Quantum Cartridge When The Quantum Cartridge should be replaced when one of the following Alert or Alarm messages is displayed

Page 61 - Placing

- 64 - Replacing the Quantum Cartridge, Continued Placing Follow the steps below to install a new Quantum Cartridge. Step Action Diagram 1 Remov

Page 62 - Cartridge

- 65 - Replacing the Quantum Cartridge, Continued Rinsing The Quantum Cartridge, when newly installed, needs to be rinsed. This ensures optimal wa

Page 63

- 66 - Replacing the Quantum Cartridge, Continued Rinsing (continued) Step Action Diagram 4 Dispense water for about 10 minutes. This flushes ou

Page 64

- 67 - Replacing a POD Pak Basing on flowrate One possible reason for a decrease in Milli-Q Water flowrate is a clogged POD Pak. The POD Pak shoul

Page 65 - Rinsing

- 68 - Replacing a POD Pak, Continued Registering (continued) Step Action Diagram 4 Scroll down to Install POD Pak. 5 Press . 6 Press . 7

Page 66

- 69 - Replacing a POD Pak, Continued Registering (continued) Step Action Diagram 9 Press . 10 Press 3 times on .

Page 68

- 70 - Cleaning the Inlet Strainer Purpose The purpose of the Inlet Strainer is to prevent a large particle from entering the System. If the Inlet

Page 69

- 71 - Cleaning the Inlet Strainer, Continued Registering (continued) Step Action Diagram 2 Select Menu. Press . 3 Select Maintenance. Press .

Page 70 - Cleaning the Inlet Strainer

- 72 - Cleaning the Inlet Strainer, Continued Registering (continued) Step Action Diagram 7 Press 3 times on . 8 Go to READY Mode.

Page 71

- 73 - Calibrating the Flowrate When The flowrate should be calibrated when a:  new consumble is installed such as a:  POD Pak, or  Q-Gard Pack,

Page 72

- 74 - Alarms Overview Introduction The purpose of this chapter is to explain the Alarm messages shown on a System. Specifically, this chapter expl

Page 73

- 75 - Alarm Information Definition An Alarm message is a way of informing you that immediate attention is required for the System. Alarm shown –

Page 74

- 76 - Alarm Information, Continued System Alarms When an Alarm is shown, it is listed under the System Alarms LCD. See the section <View Operat

Page 75 - Alarm Information

- 77 - Alarm Information, Continued Cancelling an Alarm message The display of an Alarm message can be cancelled by:  fixing the cause of the Alar

Page 76 - Alarm Information, Continued

- 78 - Alarm Information, Continued Alarm – after cancelling the text display Main Display LEDs Main Display Alarm – fixed Now suppose a Milli

Page 77

- 79 - Summary of Alarm messages Alarm messages LCD message What it means FLOW AUTO STOP The System has automatically stopped dispensing water.

Page 78

- 8 - Product Information Overview Purpose This chapter contains topics related to the System. Some of the more important topics in this chapter ar

Page 79 - Summary of Alarm messages

- 80 - Summary of Alarm messages, Continued Alarm messages (continued) LCD message What it means MILLI-Q RES > MAX The Milli-Q Water resistiv

Page 80

- 81 - Alerts Overview Introduction The purpose of this chapter is to explain the Alert messages shown on a System. Specifically, this chapter expl

Page 81

- 82 - Alert information Purpose An Alert message corresponds to a maintenance request. Most of the Alert messages are related to the replacement

Page 82 - Alert information

- 83 - Alert information, Continued Viewing an Alert Message Follow the steps below to view an Alert message. Step Action Diagram 1 Start in eit

Page 83 - Alert information, Continued

- 84 - Alert information, Continued Cancelling a Minor Alert message -procedure A Minor alert message can be cancelled by:  performing the mainte

Page 84

- 85 - Alert information, Continued Minor Alert - after cancelling The Alert message has been cancelled but the cause of the message is still activ

Page 85

- 86 - Alert information, Continued Major Alert – after cancelling The Alert message has been cancelled but the cause of the message is still activ

Page 86

- 87 - Summary of Alert messages Alert messages LCD message What it means CALIBRATION VISIT OVERDUE XX DAYS The System has determined that a Cali

Page 87 - Summary of Alert messages

- 88 - Summary of Alert messages, Continued Alert messages (continued) LCD message What it means REPLACE Q-GARD PACK The System has determined t

Page 88 - Alert messages

- 89 - Ordering Information Consumables, Accessories and Systems Consumables Item Catalogue Number BioPak Ultrafilter CDUFBI001 Millipak Express

Page 89 - Ordering Information

- 9 - Cabinet Overview Item Description/Name A Point Of Delivery (POD) B POD Pak C POD Arm D Connections for tubings, power cord, level sensor

Page 90 - Reference

- 90 - Consumables, Accessories and Systems, Continued Milli-Q Reference System Item Catalogue Number Milli-Q Reference Cabinet Z00QSV001 NOTE: A