User Manual Milli-Q Reference System
- 10 - Cabinet, Continued Details of the Main Display Item Description 1 LEDs 2 Main LCD 3 Main Keypad Right The use of the Right Keypad button
- 11 - Cabinet, Continued Up The use of the Up Keypad button is shown below. It is used to scroll up in a menu. Diagram 1 Action Diagram 2 Pres
- 12 - Cabinet, Continued READY Mode – water quality values The READY Mode screen display is explained below. Diagram Explanation In this example
- 13 - Cabinet, Continued Port and cables The port and cable connections are explained below. Item Description Item Description 1 Feedwater por
- 14 - Consumables Flow diagram The water flow through a System is shown here in a flow diagram. The various consumables are described below. Ite
- 15 - Consumables, Continued Quantum Cartridge The Quantum Cartridge removes trace levels of ions and organic molecules. Item Description Quantum
- 16 - Specifications and requirements Milli-Q® Water quality The water delivered from a POD Unit has the following characteristics. Parameter Spe
- 17 - Specifications and requirements, Continued Dimensions Materials of construction Please contact Millipore for a list of the Materia
- 18 - Specifications and requirements, Continued Feedwater The Feedwater requirements are listed here. Parameter Value Type Pre-treated water inc
- 19 - Installation Overview Purpose This chapter explains how to install the System. Contents This chapter contains the following topics: Topic
- 2 - About this User Manual Purpose This User Manual is intended for use with a Milli-Q® Reference Water Purification System. This User Manual is
- 20 - Alarms generated during installation Overview During the installation of a System, certain Alarm messages are generated. This occurs because
- 21 - POD Unit, tubing and power cord Separate POD Arm and Point Of Delivery Separate the POD Arm and the Point Of Delivery by cutting and removin
- 22 - POD Unit, tubing and power cord, Continued Loop Install the Inlet Strainer as shown here. Connect one end of the feedwater tubing to the
- 23 - POD Unit, tubing and power cord, Continued Alarm messages Because the System is starting without a Q-Gard Pack or a Quantum Cartridge instal
- 24 - Installing the Q-Gard Pack Procedure Follow the steps below to install a new Q-Gard Pack. Step Action Diagram 1 Start in STANDBY Mode. NO
- 25 - Installing the Q-Gard Pack, Continued Procedure (continued) Step Action Diagram 5 Push the bottom of the Q-Gard Pack inwards. 6 Push th
- 26 - Installing the Quantum Cartridge Procedure Follow the steps below to install a new Quantum Cartridge. Step Action Diagram 1 Open the righ
- 27 - Rinsing the System Procedure Follow the steps below to rinse the System. Step Action Diagram 1 Locate the clear tubing and the barbed fit
- 28 - Rinsing the System, Continued Procedure (continued) Step Action Diagram 5 Push the POD Plunger all the way down and then release it to st
- 29 - Installing a POD Pak Overview The installation of a POD Pak involves 2 steps. These are: placing and flushing the POD Pak onto the POD Un
- 3 - Legal Information Notice The information in this document is subject to change without notice and should not be construed as a commitment by
- 30 - Installing a POD Pak, Continued Registering (continued) Step Action Diagram 5 Press . 6 Press . 7 In this example, you choose Millipa
- 31 - Registering UV Lamp timer Introduction The timer used for the UV 185 nm Lamp must be reset when the System is installed. If this is not done
- 32 - Registering UV Lamp timer, Continued Procedure (continued) Step Action Diagram 5 Press . 6 Press . 7 Press . 8 Press 3 times on .
- 33 - Calibrating the Flowrate Introduction The Milli-Q Water flowrate should be calibrated when the System is installed. A 1 Litre graduated cyli
- 34 - Calibrating the Flowrate, Continued Procedure (continued) Step Action Diagram 6 Place a 1 L Graduated Cylinder under the POD Unit. Press
- 35 - Software Overview Introduction The purpose of this chapter is to explain the various software used in the System. Contents This chapter con
- 36 - Software Map Map
- 37 - Standby Mode General information Purpose STANDBY mode is used primarily for: maintenance actions, and going to the Manager Menu. Displa
- 38 - Description of Standby Menu Maintenance The Maintenance Menu is described below. Diagram 1 Diagram 2 Item Description Clean Strainer
- 39 - Description of Standby Menu, Continued Language Diagram 1 Diagram 2 Item Description Language Change the displayed language.
- 4 - Legal Information Continued Trademarks Millipore, Elix, Q-Gard, Quantum, Milli-Q, BioPak, EDS-Pak, Millipak and Millipore Express are registe
- 40 - Manager Menu Description How to enter See the Software Map at the beginning of this chapter. The map shows how to enter the Manager Menu. T
- 41 - Description, Continued Set Points Diagram 1 Diagram 2 Item Description Strainer Frequency Change set points for controll
- 42 - Description, Continued Setup Diagram 1 Diagram 2 Item Description Install Date Change the installation date. Buzzer Ch
- 43 - Ready Mode General information Purpose In READY Mode, water can be dispensed from the POD Unit. The System should be left in READY Mode mos
- 44 - Description of Ready Menu Water Quality Diagram 1 Diagram 2 Item Description MQ Feed Quality View the feedwater quality (accessory) T
- 45 - Description of Ready Menu, Continued Consumables Status Diagram 1 Diagram 2 Consumable Description Q-Gard UV 185 nm Lamp Q
- 46 - Description of Ready Menu, Continued Service Tracking Diagram 1 Diagram 2 Item Description Installation Repair Service Contract Contra
- 47 - Using the Milli-Q System Overview Introduction The purpose of this chapter is to explain: various ways that water can be dispensed from th
- 48 - Dispensing water Optimise Water Quality Product Water can be recirculated within the System before dispensing it. This helps optimised wate
- 49 - Dispensing water, Continued Using the POD Plunger To dispense water, press down on the POD Unit plunger while in READY Mode. Position Wa
- 5 - Safety Information Statement Your Milli-Q System should be installed and operated according to the instructions in this manual. In particular
- 50 - Dispensing water, Continued Volumetric dispensing (continued) Step Action Diagram 3 Select the desired volume of water to be delivered. P
- 51 - Viewing water quality Procedure Follow the steps below to view the water quality. Step Action Diagram 1 Make sure the System is in READY
- 52 - Viewing Operation Introduction VIEW OPERATION allows you to see the status of major components. Under the View Operation LCD, the following
- 53 - Viewing Operation, Continued System Alerts An example Alert is shown here. This is an Alert that is currently being displayed on the botto
- 54 - Viewing Consumable Status Introduction Consumables Status allows you to see information related to the various consumables. Procedure Follo
- 55 - Calling Millipore Introduction Call Millipore allows you to see contact information. A Millipore Service Representative can enter this infor
- 56 - Viewing Information Introduction INFORMATION allows you to view: flow schematic information, version information, and serial number an
- 57 - Viewing Information, Continued Version The various versions for the System are shown here. This LCD shows the version used for various compo
- 58 - Maintenance Overview Introduction The purpose of this chapter is to explain the common maintenance needed for a System. Contents This chapt
- 59 - Maintenance Schedule Consumables Item Maintenance needed When Q-Gard Pack Replacement Prompted to by an LCD message. Quantum Cartridge
- 6 - Table of Contents Product Information ...
- 60 - Replacing the Q-Gard Pack When The Q-Gard Pack should be replaced when one of the following Alarm or Alert messages is displayed. Alarm me
- 61 - Replacing the Q-Gard Pack, Continued Removing (continued) Step Action Diagram 4 Remove the used Q-Gard Pack. 5 The System will indicate
- 62 - Replacing the Q-Gard Pack, Continued Placing (continued) Step Action Diagram 3 Push the Pack Locking Handle down. Close the left door.
- 63 - Replacing the Quantum Cartridge When The Quantum Cartridge should be replaced when one of the following Alert or Alarm messages is displayed
- 64 - Replacing the Quantum Cartridge, Continued Placing Follow the steps below to install a new Quantum Cartridge. Step Action Diagram 1 Remov
- 65 - Replacing the Quantum Cartridge, Continued Rinsing The Quantum Cartridge, when newly installed, needs to be rinsed. This ensures optimal wa
- 66 - Replacing the Quantum Cartridge, Continued Rinsing (continued) Step Action Diagram 4 Dispense water for about 10 minutes. This flushes ou
- 67 - Replacing a POD Pak Basing on flowrate One possible reason for a decrease in Milli-Q Water flowrate is a clogged POD Pak. The POD Pak shoul
- 68 - Replacing a POD Pak, Continued Registering (continued) Step Action Diagram 4 Scroll down to Install POD Pak. 5 Press . 6 Press . 7
- 69 - Replacing a POD Pak, Continued Registering (continued) Step Action Diagram 9 Press . 10 Press 3 times on .
- 7 - Alarms ...
- 70 - Cleaning the Inlet Strainer Purpose The purpose of the Inlet Strainer is to prevent a large particle from entering the System. If the Inlet
- 71 - Cleaning the Inlet Strainer, Continued Registering (continued) Step Action Diagram 2 Select Menu. Press . 3 Select Maintenance. Press .
- 72 - Cleaning the Inlet Strainer, Continued Registering (continued) Step Action Diagram 7 Press 3 times on . 8 Go to READY Mode.
- 73 - Calibrating the Flowrate When The flowrate should be calibrated when a: new consumble is installed such as a: POD Pak, or Q-Gard Pack,
- 74 - Alarms Overview Introduction The purpose of this chapter is to explain the Alarm messages shown on a System. Specifically, this chapter expl
- 75 - Alarm Information Definition An Alarm message is a way of informing you that immediate attention is required for the System. Alarm shown –
- 76 - Alarm Information, Continued System Alarms When an Alarm is shown, it is listed under the System Alarms LCD. See the section <View Operat
- 77 - Alarm Information, Continued Cancelling an Alarm message The display of an Alarm message can be cancelled by: fixing the cause of the Alar
- 78 - Alarm Information, Continued Alarm – after cancelling the text display Main Display LEDs Main Display Alarm – fixed Now suppose a Milli
- 79 - Summary of Alarm messages Alarm messages LCD message What it means FLOW AUTO STOP The System has automatically stopped dispensing water.
- 8 - Product Information Overview Purpose This chapter contains topics related to the System. Some of the more important topics in this chapter ar
- 80 - Summary of Alarm messages, Continued Alarm messages (continued) LCD message What it means MILLI-Q RES > MAX The Milli-Q Water resistiv
- 81 - Alerts Overview Introduction The purpose of this chapter is to explain the Alert messages shown on a System. Specifically, this chapter expl
- 82 - Alert information Purpose An Alert message corresponds to a maintenance request. Most of the Alert messages are related to the replacement
- 83 - Alert information, Continued Viewing an Alert Message Follow the steps below to view an Alert message. Step Action Diagram 1 Start in eit
- 84 - Alert information, Continued Cancelling a Minor Alert message -procedure A Minor alert message can be cancelled by: performing the mainte
- 85 - Alert information, Continued Minor Alert - after cancelling The Alert message has been cancelled but the cause of the message is still activ
- 86 - Alert information, Continued Major Alert – after cancelling The Alert message has been cancelled but the cause of the message is still activ
- 87 - Summary of Alert messages Alert messages LCD message What it means CALIBRATION VISIT OVERDUE XX DAYS The System has determined that a Cali
- 88 - Summary of Alert messages, Continued Alert messages (continued) LCD message What it means REPLACE Q-GARD PACK The System has determined t
- 89 - Ordering Information Consumables, Accessories and Systems Consumables Item Catalogue Number BioPak Ultrafilter CDUFBI001 Millipak Express
- 9 - Cabinet Overview Item Description/Name A Point Of Delivery (POD) B POD Pak C POD Arm D Connections for tubings, power cord, level sensor
- 90 - Consumables, Accessories and Systems, Continued Milli-Q Reference System Item Catalogue Number Milli-Q Reference Cabinet Z00QSV001 NOTE: A