iilli-Q* PlusWater Purification SystemInstallation and OperationM1LUPORE
B. Installation1. UnpackingUnpack the system carefully from its container.Verify Contents: 1. Milli-Q Plus system2. Operation and maintenance manual3.
B. Installation2. Battery ActivationBefore making water, or electrical connections (initial startup)the battery, to power the system memory, must be a
Note: Switch on power immediately after battery activation.Switch system to ON for a few seconds. This willinitialise the microprocessor and avoid dis
The system has a single "quick connect"connection located at the rear of thesystem.Simply push-in tube to attachTube is secured in positionP
Isolating ValveA valve should be installed on the inlet water line tofacilitate maintenance of the Milli-Q.Tank feed0,1mmin.Distribution Loop FeedRec
B. Installation4. Electrical ConnectionON/OFF SwitchFuse holderPower InletMain Connection: 110-120 V/60 Hz (0.8 A) (System Cat. No. ZD52 115 84)Ground
B. Installation5. Bench/Wall MountingThe system is compact, and hasfour rubber feet for easyinstallation on the laboratorybench.Alternatively, the sys
B. Installation6. Pack InstallationRemove plugs from pack.Remove plugs from packreceptor inside Milli-Q.Unlock pack retaining armon Milli-Q by pushing
B. Installation7. Initial Start UpChecklist: - battery in place and activated- water connected to rear of unit- power cable plugged into rear socket-
Allow system to run for 5 minutesto vessel.Close point-of-use valve. Turnto RECIRCULATION.Allow system to recirculatefor 5 minutes.Open point-of-use v
MILLI-Q PLUSA. Introduction1. System Description2. External Features3. Schematic4. Dimensions5. Function of Components/ControlsB. Installation1. Unpac
Allow water to purge air fromMillipak.Close Millipak vent by replacingcap.Close point-of-use valve. Turnto RECIRCULATION.Press OPERATE/STANDBY to go i
P0WSRI •mum®.—"OPERATE 'STANDBYrC. Operation1. Auto TestWhen the operate button ispressed, after a standbymode, the unit performs anautomati
4. Auto Recirculation (Intermittent)During STANDBY the system will automatically recirculate waterfor a period of 5 minutes every one hour (or 3 hours
7. Remote StandbyWhen the system is connected toa remote controlling unit, viathe inlet jack plug at the rearof the unit, the display willshow REMOTE
Dispose of old pack.Remove caps on new pack.Moisten 0-rings.Maintain Milli-Q Plus in position with left hand. Slide pack intoposition with right hand.
Open point-of-use valve to PRODUCTION position.Allow water to purge air from vent.Close the vent.Close point-of-use valve switch to RECIRCULATION.Pres
D. Maintenance1. Access to PC Board(Refer to Section B., Installation)Remove rear cabinet by unlocking the cabinet locking nut at rear.Disconnect 5 tr
3. Fuse ReplacementIf power lamp is off, check fuses as follows:Main ON/OFF switch, rearof unit, and socket onring main.Check main fuse. Removemain ca
5. Pack Replacement Frequency SettingThe EXCHANGE PACK light will begin to flash after the pack hasbeen in place for longer than a preset time.The pac
7. TroubleshootingFault: Initial AUTO TEST:Resistivity is not 15 Megohm x cm ± 1 Megohm x cmWhen the OPERATE/STANDBY button is first pressed, the unit
Milli-Q PlusA. INTRODUCTION1. System DescriptionThe Milli-Q Plus system is a state of the art water purificationsystem, designed to provide the final
To Recalibrate TemperatureRemove rear cabinet.Remove Dumper C (this locks PCboard to TEST mode).Press temperature button andmaintain.Display shows the
Check the TEST TEMPERATURE as previously described for TestTemperature Calibration.Readjust to 25° C if incorrect.Check if the temperature of your fee
E. Technical Information1. ComponentsComponentsCabinets front-rearInternal skeletonHardware (screws, nuts, washers)3-way valves0-ringsMicrosensor tips
2. SpecificationsDimensionsHeight 495 mmWidth 297 mmDepth 433 mmConnectionsFeed connector Polyamide tubing 8 mm O.D.quick connect fittingElectrical Re
3. Ordering InformationDescription Catalogue NumberMilli-Q Plus system 220-240V/50 Hz ZFMQ 050 01Milli-Q Plus system 110-120V/60 Hz ZD52 115 84Expenda
4. WarrantyMillipore Corporation (the "Company") warrants its Waterpurification products against defects in materials andworkmanship when us
Milli-Q* PlusReagent GradeWater Purification SystemCatalogue NumbersZFMQ 05001 (230 V/50 Hz)ZD5211584(115V/60Hz)Components and Spare PartsMILLIPORE
— LEFT SIDE VIEW1 ) ficfl (S) (5LIST OF COMPONENTSITEM DESCRIPTIONCATALOGUE NO.MILLI-Q PLUS220/240 V - 50 HzCAT. NO. 2FMQ 050 01MILLI-Q PLUS115 V/60 H
MILLIPOREr1FRONT VIEW
MILLIPOREFor more informationplease call the nearestMillipore subsidiaryNorth AmericaCall toll-free 800-225-1380In Western States 800-632-2708In Canad
A. Introduction2. External Features5 3 111 8 9 106 713 15 14 12Front View (left)1. Alphanumeric display2. Control keypad3. Pack lock/unlock arm4. Puri
PLEASE NOTEPRODUCT IMPROVEMENTMilli-Q Plus Milli-Q UF Plus Alpha-QMilli-RO 6 Milli-RO 6 Plus Milli-RO 10 Milli-RO 10 PlusAll units are now shipped wit
A. Introduction3. SchematicFlow Schematic1 - Inlet solenoid valve2 - Pump3 - Purification pack4 - Resistivity sensor5 - 3-way valve (manual)6 - Check
A. Introduction5. Function of Components/ControlsOn/Off Power SwitchThis switch, located on the rear side of the unit, controlsthe power to the contro
If this resistivity is less than the set point of 14 Megohm x cmthen the display of product , _ Megohm x cm will flash on andoff. When the resistivity
ALPHANUMERIC DISPLAYKeyboardPress OPERATE/STANDBYModeOperatePress OPERATE/STANDBY StandbyAutomatic recirculation 5 minutes Standbyevery 55 minutesPres
PURIFICATION PACKThe purification pack utilizes a chain of purificationmedia to polish the feed water to high quality:QPAK1 (to be used if feed water