iilli-Q* Plus
Water Purification System
Installation and Operation
M1LUPORE
Vue de la page 0
1 2 3 4 5 6 ... 39 40

Résumé du contenu

Page 1 - M1LUPORE

iilli-Q* PlusWater Purification SystemInstallation and OperationM1LUPORE

Page 2 - MILLI-Q PLUS

B. Installation1. UnpackingUnpack the system carefully from its container.Verify Contents: 1. Milli-Q Plus system2. Operation and maintenance manual3.

Page 3

B. Installation2. Battery ActivationBefore making water, or electrical connections (initial startup)the battery, to power the system memory, must be a

Page 4 - 13 15 14 12

Note: Switch on power immediately after battery activation.Switch system to ON for a few seconds. This willinitialise the microprocessor and avoid dis

Page 5

The system has a single "quick connect"connection located at the rear of thesystem.Simply push-in tube to attachTube is secured in positionP

Page 6

Isolating ValveA valve should be installed on the inlet water line tofacilitate maintenance of the Milli-Q.Tank feed0,1mmin.Distribution Loop FeedRec

Page 7

B. Installation4. Electrical ConnectionON/OFF SwitchFuse holderPower InletMain Connection: 110-120 V/60 Hz (0.8 A) (System Cat. No. ZD52 115 84)Ground

Page 8 - ALPHANUMERIC DISPLAY

B. Installation5. Bench/Wall MountingThe system is compact, and hasfour rubber feet for easyinstallation on the laboratorybench.Alternatively, the sys

Page 9 - Activated

B. Installation6. Pack InstallationRemove plugs from pack.Remove plugs from packreceptor inside Milli-Q.Unlock pack retaining armon Milli-Q by pushing

Page 10

B. Installation7. Initial Start UpChecklist: - battery in place and activated- water connected to rear of unit- power cable plugged into rear socket-

Page 11

Allow system to run for 5 minutesto vessel.Close point-of-use valve. Turnto RECIRCULATION.Allow system to recirculatefor 5 minutes.Open point-of-use v

Page 12

MILLI-Q PLUSA. Introduction1. System Description2. External Features3. Schematic4. Dimensions5. Function of Components/ControlsB. Installation1. Unpac

Page 13

Allow water to purge air fromMillipak.Close Millipak vent by replacingcap.Close point-of-use valve. Turnto RECIRCULATION.Press OPERATE/STANDBY to go i

Page 14 - Regulator

P0WSRI •mum®.—"OPERATE 'STANDBYrC. Operation1. Auto TestWhen the operate button ispressed, after a standbymode, the unit performs anautomati

Page 15

4. Auto Recirculation (Intermittent)During STANDBY the system will automatically recirculate waterfor a period of 5 minutes every one hour (or 3 hours

Page 16

7. Remote StandbyWhen the system is connected toa remote controlling unit, viathe inlet jack plug at the rearof the unit, the display willshow REMOTE

Page 17

Dispose of old pack.Remove caps on new pack.Moisten 0-rings.Maintain Milli-Q Plus in position with left hand. Slide pack intoposition with right hand.

Page 18

Open point-of-use valve to PRODUCTION position.Allow water to purge air from vent.Close the vent.Close point-of-use valve switch to RECIRCULATION.Pres

Page 19

D. Maintenance1. Access to PC Board(Refer to Section B., Installation)Remove rear cabinet by unlocking the cabinet locking nut at rear.Disconnect 5 tr

Page 20

3. Fuse ReplacementIf power lamp is off, check fuses as follows:Main ON/OFF switch, rearof unit, and socket onring main.Check main fuse. Removemain ca

Page 21

5. Pack Replacement Frequency SettingThe EXCHANGE PACK light will begin to flash after the pack hasbeen in place for longer than a preset time.The pac

Page 22

7. TroubleshootingFault: Initial AUTO TEST:Resistivity is not 15 Megohm x cm ± 1 Megohm x cmWhen the OPERATE/STANDBY button is first pressed, the unit

Page 23

Milli-Q PlusA. INTRODUCTION1. System DescriptionThe Milli-Q Plus system is a state of the art water purificationsystem, designed to provide the final

Page 24

To Recalibrate TemperatureRemove rear cabinet.Remove Dumper C (this locks PCboard to TEST mode).Press temperature button andmaintain.Display shows the

Page 25

Check the TEST TEMPERATURE as previously described for TestTemperature Calibration.Readjust to 25° C if incorrect.Check if the temperature of your fee

Page 26

E. Technical Information1. ComponentsComponentsCabinets front-rearInternal skeletonHardware (screws, nuts, washers)3-way valves0-ringsMicrosensor tips

Page 27 - # *--—"

2. SpecificationsDimensionsHeight 495 mmWidth 297 mmDepth 433 mmConnectionsFeed connector Polyamide tubing 8 mm O.D.quick connect fittingElectrical Re

Page 28

3. Ordering InformationDescription Catalogue NumberMilli-Q Plus system 220-240V/50 Hz ZFMQ 050 01Milli-Q Plus system 110-120V/60 Hz ZD52 115 84Expenda

Page 29

4. WarrantyMillipore Corporation (the "Company") warrants its Waterpurification products against defects in materials andworkmanship when us

Page 30

Milli-Q* PlusReagent GradeWater Purification SystemCatalogue NumbersZFMQ 05001 (230 V/50 Hz)ZD5211584(115V/60Hz)Components and Spare PartsMILLIPORE

Page 31

— LEFT SIDE VIEW1 ) ficfl (S) (5LIST OF COMPONENTSITEM DESCRIPTIONCATALOGUE NO.MILLI-Q PLUS220/240 V - 50 HzCAT. NO. 2FMQ 050 01MILLI-Q PLUS115 V/60 H

Page 32

MILLIPOREr1FRONT VIEW

Page 33

MILLIPOREFor more informationplease call the nearestMillipore subsidiaryNorth AmericaCall toll-free 800-225-1380In Western States 800-632-2708In Canad

Page 34

A. Introduction2. External Features5 3 111 8 9 106 713 15 14 12Front View (left)1. Alphanumeric display2. Control keypad3. Pack lock/unlock arm4. Puri

Page 35

PLEASE NOTEPRODUCT IMPROVEMENTMilli-Q Plus Milli-Q UF Plus Alpha-QMilli-RO 6 Milli-RO 6 Plus Milli-RO 10 Milli-RO 10 PlusAll units are now shipped wit

Page 36 - Milli-Q*

A. Introduction3. SchematicFlow Schematic1 - Inlet solenoid valve2 - Pump3 - Purification pack4 - Resistivity sensor5 - 3-way valve (manual)6 - Check

Page 37 - COMPONENTS

A. Introduction5. Function of Components/ControlsOn/Off Power SwitchThis switch, located on the rear side of the unit, controlsthe power to the contro

Page 38 - FRONT VIEW

If this resistivity is less than the set point of 14 Megohm x cmthen the display of product , _ Megohm x cm will flash on andoff. When the resistivity

Page 39 - MILLIPORE

ALPHANUMERIC DISPLAYKeyboardPress OPERATE/STANDBYModeOperatePress OPERATE/STANDBY StandbyAutomatic recirculation 5 minutes Standbyevery 55 minutesPres

Page 40 - PLEASE NOTE

PURIFICATION PACKThe purification pack utilizes a chain of purificationmedia to polish the feed water to high quality:QPAK1 (to be used if feed water