PRODUCT INFORMATION
Milli-Q Academic / Academic A10 12
2-4 TECHNICAL SPECIFICATIONS
DIMENSIONS (SYSTEM AND CLEARANCE)
The height, width and depth of a Milli-Q Academic (and Academic A10) are shown here:
WEIGHT
Milli-Q Academic: Operating weight (wetted Q-Gard, Quantum and Millipak): 16 kg.
Milli-Q Academic A10: Operating weight: 16.8 kg.
ELECTRICAL
The electrical requirements for a Milli-Q Academic (or Milli-Q Academic A10) are:
120 VAC 10%, 50/60 Hz, 60 VA, 0.5 amp source, 2 amp Time Lag fuse used (ref. ZMQS600T1,
ZMQP600T1, ZMQS60001, ZMQP60001).
230 VAC 10%, 50/60 Hz, 60 VA, 0.26 amp source, 1 amp Time Lag fuse used (ref. ZMQS500T1,
ZMQP500T1, ZMQS50001, ZMQP50001).
The source of electrical power should be within 1 meter of the Milli-Q. The source of electrical power must be
earth grounded.
MEASUREMENT RANGES FOR TOC, RESISTIVITY, CONDUCTIVITY AND TEMPERATURE
The Milli-Q can display various Product Water quality parameters. The range of measurement of each parameter is
written below.
Resistivity Measurement Range: 2.4 – 18.2 M.cm (when compensated to a 25C value)
Conductivity Measurement Range: 0.055 – 0.294 S/cm
Temperature Measurement Range: 0 – 45 C (not the system temperature operating range)
TOC Measurement Range: 1 – 999 ppb
256 mm
296 mm
445 mm
74 mm
778 mm
336 mm
20 mm
190 mm