INSTALLATION
Milli-Q Academic / Academic A10 32
4-8 HYDRATING THE SYSTEM
At this time, the Millipak (or BioPak) is not installed. The Barbed Fitting and Tygon Tubing should still be
attached to the end of the POU Dispenser.
Leave the Milli-Q in PRE OPERATE Mode overnight or for several hours. This will aid in rinsing off the
purification media inside the Q-Gard Pack and the Quantum Cartridge. Do not leave the Milli-Q in STANDBY
Mode.
(The next morning) Bring the POU Dispenser Trigger forward to put the Milli-Q into PRODUCT Mode.
Allow about 10 to 15 Litres (about 7 to 10 minutes) of water to be dispensed from the Milli-Q.
Bring the POU Dispenser Trigger upright. The Milli-Q will finish dispensing water. The Milli-Q will continue to
operate for 90 seconds after which it will go into PRE OPERATE Mode.
Leave the Milli-Q in PRE OPERATE Mode.
4-9 INSTALLATION AND RINSING OF THE MILLIPAK
Obtain a Millipak.
Remove the Tygon Tubing and the Barbed Fitting from the POU Dispenser.
Screw the Millipak onto the POU Dispenser. Do not use any Teflon Tape on the threads of the
Millipak. The Millipak should be turned until it is handtight. Do not overtighten the Millipak.
Do not use Teflon Tape on the threads of the Millipak.
The POU
Dispenser has an O-ring inside which provides a watertight seal.
Move the POU Dispenser Trigger forwards. This will bring the Milli-Q System into PRODUCT Mode.
Locate the air vent valve on the top side (clear side) of the Millipak. Open this slowly but do not remove it
from the Millipak. Allow any air on the clear side of the Millipak to be vented out.
Dispense about 5 Litres of water. Make sure all air is purged out.
Air will not pass through the membrane filter in the Millipak. If
there is trapped air in the Millipak, then a reduction in Product Water flowrate
can result.
Move the POU Dispenser Trigger back to the upright position. The Milli-Q should now be in PRE OPERATE
Mode 90 seconds later. Leave the Milli-Q in PRE OPERATE Mode.
It is highly recommended not to put Tygon Tubing or any other
type of tubing on the end of the Millipak. This can compromise the
Product Water quality (bacteria can grow in the Tygon Tubing).
The Millipak is now installed and rinsed out.