INSTALLATION
Milli-Q Academic / Academic A10 31
The LCD will change and will look like:
Let the AIR PURGE continue for its 5 minute duration. It will count down to 0 minutes (5, 4,..., 0 mn).
Pressurised Feedwater only:
Check the Pressure Gauge on the Pressure Regulator. While water is
flowing through it, the Pressure Gauge should show 0.3 bar. If it is not set to 0.3 bar, then turn the round
handle on the Pressure Regulator clockwise until the pressure gauge (on the regulator) shows 0.3 bar. Check
this again when water is flowing through the Pressure Regulator
.
The setting of the Pressure Regulator should not exceed 0.3 bar.
The feedwater pressure reaching the Milli-Q in PRODUCT Mode
should not exceed 0.3 bar. The Pressure Regulator should only be
set while water is flowing through it. It is perfectly normal to see
an increase in the Pressure Regulator setting while water is not
flowing through it (i.e. STANDBY or PRE OPERATE Mode).
The Milli-Q will now go to STANDBY Mode and will remind you to close the POU Dispenser Valve.
Move the POU Dispenser Trigger back (its upright position). The Milli-Q will now go to PRE OPERATE Mode.
Milli-Q Academic A10 only: When you ‘Close The Valve’, the Milli-Q will go to PRE
OPERATE Mode and then will start an automatic 60 minute A10 CLEANING.
The A10 CLEANING
is a way of flushing out any particles that may have been present in the newly installed Q-Gard Pack or
Quantum Cartridge.
Milli-Q Academic A10 only:
Allow the A10 CLEANING to count down to 0 minutes
.
The remaining steps of this section apply to both the Milli-Q Academic and the Milli-Q Academic A10.
Any remaining air in Quantum Cartridge can be ‘hand purged’ out now.
Get a screwdriver with a very thin shaft and small head.
Locate the small hole on the top part of the Quantum Cartridge Door.
Gently insert the screwdriver into the Quantum Cartridge. This will open a small vent that will expel any air in
the Quantum Cartridge.
Be very careful when pushing the tip of the screwdriver into the
Vent Hole of the Quantum Cartridge. The Vent Hole only needs to
be opened a very small amount.
If all of the air does not expel out, do not worry. You can repeat this step later after more water has flowed
through the Milli-Q. Remember to have the Milli-Q in STANDBY or PRE OPERATE Mode while you are
expelling air out of the Quantum Cartridge.
PRODUCT
AIR PURGE 5 mn
P R E O P E R A T E
O P E N T H E V A L V E
ST A ND B Y
C L O S E T H E V A L V E
PRE OPERATE
S T A ND B Y
C L O S E T H E V A L V E
ATTENTION
ATTENTION
PRE OPERATE
A10 CLEANING: 60
S T A ND B Y
C L O S E T H E V A L V E