USING THE MILLI-Q
Milli-Q Academic / Academic A10 53
5-7 HOW TO UNDERSTAND MILLI-Q MESSAGES
It is possible for various messages to be displayed on the Milli-Q LCD while you are using the Milli-Q. While a
message is being displayed, either the Yellow Service LED or the Red Alarm LED will be blinking.
The table below provides a summary of these messages, what they mean and what to do when you see them.
You may also refer to Chapter 7 (Troubleshooting) for more information about these messages and
other troubleshooting items.
Please contact Millipore if you desire more information regarding what to do when a message appears on the
Milli-Q.
Message and
System Status
What it means What to do
EXCH. CARTRIDGES
Milli-Q Academic: The Q-Gard
and Quantum have been installed for
more than 180 Calendar Days or the
Resistivity is below the Resistivity
Setpoint (Factory Value is
16.0 M.cm @ 25C).
Replace the Q-Gard and the
Quantum (and the Millipak or
BioPak).
EXCH. CARTRIDGES
Milli-Q Academic A10: The
Resistivity is below the Resistivity
Setpoint (Factory Value is
16.0 M.cm @ 25C).
Replace the Q-Gard and the
Quantum (and the Millipak or
BioPak).
TOC > SETPOINT
The displayed TOC (Milli-Q
Academic A10 only) is greater than
the TOC Setpoint established in the
software.
A TOC value above the TOC
Setpoint can have many causes.
Contact Millipore.
CARTRIDGE OUT
The Milli-Q will not operate. This
message must be resolved before
the Milli-Q can operate.
Either the Q-Gard Pack or the
Quantum Cartridge is not installed
correctly or one of them is absent.
Reinstall either or both the
Q-Gard Pack or the Quantum
Cartridge.
NO FEED WATER
The Milli-Q will not operate. This
message must be resolved before
the Milli-Q can operate.
An accessory has detected that there
is no feedwater going to the Milli-Q.
This accessory is not factory installed
and needs to be installed in the
system.
Make sure the Reservoir or the
Pressurised Pipe feeding the
Milli-Q has a supply of
feedwater that can be delivered
to the Milli-Q.