USING THE MILLI-Q
Milli-Q Academic / Academic A10 55
Message and
System Status
What it means What to do
A10 ERROR 0
A10 PCB E
2
Prom defective. Turn off/on the Milli-Q and dispense
water for > 6 minutes. If the
message continues, then contact
Millipore.
A10 ERROR 1
A10 PCB A/D converter defective. Turn off/on the Milli-Q and dispense
water for > 6 minutes. If the
message continues, then contact
Millipore.
A10 ERROR 2
A10 Thermistor defective. Turn off/on the Milli-Q and dispense
water for > 6 minutes. If the
message continues, then contact
Millipore.
A10 ERROR 3
Problem occurred with temperature
compensation.
Turn off/on the Milli-Q and dispense
water for > 6 minutes. If the
message continues, then contact
Millipore.
A10 ERROR 4
The water entering the A10 is < 4C. Contact Millipore.
A10 ERROR 5
The water entering the A10 is >
41C.
Contact Millipore.
A10 ERROR 6
The conductivity of the water entering
the A10 is > 1.1 S/cm.
Contact Millipore.
A10 ERROR 7
The temperature inside the A10 during
its analysis mode reached > 55 C.
Contact Millipore.
A10 ERROR 9
The TOC is very low or a problem
occurred with the A10.
Contact Millipore.